DVI Mode Operation
The DVI mode provides dual chamber AV sequential pacing without atrial
sensing/tracking. Atrial pacing occurs at the Lower Rate, with ventricular
pacing at the end of the PAV interval unless inhibited. Sensing occurs only
in the ventricle, and intrinsic atrial events are ignored. A ventricular
nonrefractory sensed event during the VA interval starts a new VA interval.
Example of DVI Mode Operation
Pacing Modes 10-19
V
P
V
S
V
S
A
P
Lower Rate Interval
Parameters:
Lower Rate = 60 ppm (1000 ms)
200 ms
A
P
Lower Rate
VA Interval
A
P
V
P
PAV Interval = 200 ms